Using the Osmometer K-7400
11
°C
10
0
-10
0 1 2 3
min
Freezing point
T
Super-cooling
Fig. 7
Temperature-time-curve for a sample solution
During a measurement the actual temperature is displayed. The
instrument detects the maximum temperature reached, stores it, and if
crystallization occurred satisfactory, displays the result.
After finishing the measurement the sample vessel is warmed up to 20°C.
As this is the stand-by temperature of the cooling cavity, the measuring
head can remain there before and after measurement as long as desired.
Using the Osmometer K-7400
Switching on
After switching ON the display presents for a short time information
concerning the instrument and the firmware version:
KNAUER
V01.XX
Freezing Point Osmometer
Now the osmometer is ready for use. In the display the main menu is
shown with the result of the last measurement. The cursor is placed on
the rhombus field
♦
.
>>> 1234 mOsmol <<<
♦
<next
sample>
Depending on the actual settings the measurement value can also be
shown as a percentage or temperature value.
>>> 12.34 % <<<
♦
<next
sample>
>>> -1.234 °C <<<
♦
<next
sample>
Содержание K-7400