Knauer HPLC IFU 2.1 Скачать руководство пользователя страница 5

6

Overview and Connectors

IFU 2.1 User Manual V6805, Version 2.0

  

The red LED „Error“ lights up if a calibration error is detected while running the 
self-test or if the software sends a signal that is not recognized by the device.

Analog Inputs (AN.IN)

Serving the purpose of receiving analog signals, up to 4 channels or devices can 
be connected to the interface box. Channels are independent of each other and 
can be used simultaneously by one system or various systems. 

 

Analog Outputs (AN.OUT)

These outputs send analog voltage signals to control devices, which have the 
necessary inpus. According to the number of outputs, four of these devices can 
be controlled with one interface box. The outputs can be programmed inde-
pendently of each other. This e. g. refers to the control of pumps in a high pres-
sure gradient system.

 

Digital Inputs (DIG.IN)

These four inputs are trigger inputs receiving a start signal from the injection 
system. They work with contact end on ground and are compatible for TTL/
CMOS or open collectors. If the channels are used for different systems, it is 
possible to start each channel individually.

 

Fig. 3 LED

Voltage range

2.56 to +2.56 V (bipolar mode)

Absolute maximum ratings

10 V to +10 V

Input impedance

10 MΩ

Maximum resolution

24 bit

Minimum noise level

7 μV (1 Hz, time constant 0.1 s)

Maximum data rate

up to 10 Hz (each channel)

Gain factor (for all channels)

1, 2, 4, 8, 16

Voltage range

0 V to +10 V

Minimum voltage step (DAC resolu-
tion)

2.5 mV (12 bit)

Maximum load resistance

2 kΩ

Max. high level input voltage

+15 V

Min. high level input voltage

+2.5 V

Содержание HPLC IFU 2.1

Страница 1: ...Interface Box IFU 2 1 User Manual V6805 HPLC...

Страница 2: ...verview and Connectors 5 Front View of the Device 5 LED 5 Analog Inputs AN IN 6 Analog Outputs AN OUT 6 Digital Inputs DIG IN 6 Digital Outputs DIG OUT 7 Installation 7 Power Supply 7 Connecting the I...

Страница 3: ...ther informations comes included with the manual of the device The interface box is no standalone device which means that software control is strictly required e g KNAUER ChromGate or KNAUER ClarityCh...

Страница 4: ...tion The green LED Power shows that the device has successfully been connected to a computer The yellow LED Busy shortly lights up during calibration or after receiving a signal from the control softw...

Страница 5: ...The outputs can be programmed inde pendently of each other This e g refers to the control of pumps in a high pres sure gradient system Digital Inputs DIG IN These four inputs are trigger inputs recei...

Страница 6: ...easures Use the device according to the specifications Connecting the Interface Box to a Computer To establish a connection between the interface box and the computer you take the USB cable from the a...

Страница 7: ...user manual of the device Please pay attention to the allowed maximum voltage Otherwise the device can get damaged KNAUER can not be held reliable in this case The required components of the remote co...

Страница 8: ...g input of a channel can not be used at the same time except the analog input is used for autozero of a detector Process Figure 1 Guide the blunt side of the depres sor tool into the square opening of...

Страница 9: ...10 Technical Data IFU 2 1 User Manual V6805 Version 2 0 Technical Data Fig 5 Exemplary cable plan Power connection via USB 2 0 5 V max 500 mA Dimensions 105 x 26 x 161 5 mm W H D Weight 0 35 kg...

Страница 10: ...shall expire in the event that any unauthorized changes are made to the device This warranty also excludes the following accidental or willful damage damage or errors caused by third parties that are...

Страница 11: ...version 2 0 2016 04 20 Printed in Germany on environmentally friendly paper from sustainable forests See up to date manuals online www knauer net downloads HPLC SMB Osmometry KNAUER Wissenschaftliche...

Отзывы: