8

Installation

IFU 2.1 User Manual V6805, Version 2.0

OpenLAB

®

 and ChromGate®: Upon registration you are asked to enter the 

serial number of the interface box of which you need to enter the last 5 digits 
only. 

EuroChrom®: When the software requests the serial number of the interface 
box in the hardware module, only the last 5 digits can be seen.

Analog Inputs

If you are to connect the analog output of a device (e. g. 1 V integrator output 
of a detector) with the analog input of the interface box, use a RCA cable sup-
plied or recommended by the manufacturer. Alternatively use a matching cable 
included with the interface box accessory kit. Please pay attention to the 
allowed maximum voltage. Otherwise the device can get damaged. 

Note: 

The analog output and digital input of the same channel cannot be used 

at the same time.

Analog Outputs

The analog outputs have to match the control input of the connected device. 
Information on the control inputs come included with the user manual of the 
device. Please pay attention to the allowed maximum voltage. Otherwise the 
device can get damaged. KNAUER can not be held reliable in this case. The 
required components of the remote connector and the USB cable are part of 
the accessories kit of the interface box. 

Note: 

The analog input and digital output of the same channel cannot be used 

at the same time.

Digital Connections

Digital inputs can be used to forward trigger signals coming from the injection 
unit. The connector for the trigger signal is the digital input of the channel, 
which also servers for recording data from the detector.

Digital outputs can be programmed via control software. 

The required components of the remote connector and the USB cable are part 
of the accessories kit of the interface box. If you require a NO connect the con-
tacts 2 and 3 (see rear view) or if you require NC connect the contacts 1 and 2 
(see rear view).

Note: 

If the analog input of a channel is used, it is not possible to program the 

digital output of the same channel. In this case, the digital output D.OUT sends 
a sampled signal at the start of run that e. g. can be used as autozero signal for 
a detector.

Содержание HPLC IFU 2.1

Страница 1: ...Interface Box IFU 2 1 User Manual V6805 HPLC...

Страница 2: ...verview and Connectors 5 Front View of the Device 5 LED 5 Analog Inputs AN IN 6 Analog Outputs AN OUT 6 Digital Inputs DIG IN 6 Digital Outputs DIG OUT 7 Installation 7 Power Supply 7 Connecting the I...

Страница 3: ...ther informations comes included with the manual of the device The interface box is no standalone device which means that software control is strictly required e g KNAUER ChromGate or KNAUER ClarityCh...

Страница 4: ...tion The green LED Power shows that the device has successfully been connected to a computer The yellow LED Busy shortly lights up during calibration or after receiving a signal from the control softw...

Страница 5: ...The outputs can be programmed inde pendently of each other This e g refers to the control of pumps in a high pres sure gradient system Digital Inputs DIG IN These four inputs are trigger inputs recei...

Страница 6: ...easures Use the device according to the specifications Connecting the Interface Box to a Computer To establish a connection between the interface box and the computer you take the USB cable from the a...

Страница 7: ...user manual of the device Please pay attention to the allowed maximum voltage Otherwise the device can get damaged KNAUER can not be held reliable in this case The required components of the remote co...

Страница 8: ...g input of a channel can not be used at the same time except the analog input is used for autozero of a detector Process Figure 1 Guide the blunt side of the depres sor tool into the square opening of...

Страница 9: ...10 Technical Data IFU 2 1 User Manual V6805 Version 2 0 Technical Data Fig 5 Exemplary cable plan Power connection via USB 2 0 5 V max 500 mA Dimensions 105 x 26 x 161 5 mm W H D Weight 0 35 kg...

Страница 10: ...shall expire in the event that any unauthorized changes are made to the device This warranty also excludes the following accidental or willful damage damage or errors caused by third parties that are...

Страница 11: ...version 2 0 2016 04 20 Printed in Germany on environmentally friendly paper from sustainable forests See up to date manuals online www knauer net downloads HPLC SMB Osmometry KNAUER Wissenschaftliche...

Отзывы: