BlueShadow Detector 40D Instructions V7647
11. Technical data
11.1 Main features
Detector type
Variable single wavelength UV/VIS
detector
Variable single wavelength UV/VIS
detector with fiber optic connectors
Detection channels
1
Light source
Deuterium (D2) lamp with integrated GLP
chip
Wavelength range
190–750 nm
Optical bandwidth
11 nm at H
α
line (FWHM)
Wavelength accu racy
± 2 .5 nm
Wavelength preci sion
0 .3 nm (ASTM E1657-98)
Noise
± 1 .5 × 10
-5
AU at 254 nm
± 2 .0 x 10
-5
AU at 254 nm
(fiber optics version)
(ASTM E1657-98)
Drift
3 .0 × 10
-4
AU/h at 254 nm
4 .0 x 10
-4
AU/h at 254 nm
(fiber optics version)
(ASTM E1657-98)
Linearity
> 2 .0 AU at 270 nm (ASTM E1657-98)
Time constants
0 .1/ 0 .2/ 0 .5/0 .1/ 1 .0/ 2 .0/5 .0/ 10 .0 s
Integration time
Automatic
11.2 Communication
Maximum data rate
50 Hz (LAN)
20 Hz (Analog)
10 Hz (RS-232)
Inputs
Error (IN)
Start (IN)
Autozero
0–10 V Analog IN
Outputs
Events 1–3
+5 V
24 V Valve