40
Operation
BlueShadow Detector 40D Instructions V7647
5.6.6 Creating and running a program with preselec-
table start time (Wake-Up Program)
The detector can start a program at a preset time (wake-up program) .
This program is preset at number 20 . The detector starts the selected
program at the specified time and date . The display switches from stand-
by (“Waiting for WakeUp”) to “R” (Run) mode .
Process
Figure
1.
Select the Device Programs menu .
2.
Enter program number 20 .
3.
Enter wake-up time, -date and the program
number you wish to start by navigating via the
horizontal arrow keys
and editing via the
numeric pad after pressing the confirmation
key
or using the vertical arrow keys
.
4.
To run the wake-up program, follow the pro-
cedure as if running any other program and
confirm the start of the wake-up program 20 .
5.
Upon choosing program number 20 the de-
tector will go into standby mode (status LED
switches to blue) until the set wake-up time and
date .
6.
Before running the program, the detector will
perform a self-calibration .
7.
By pressing the standby button for 2 seconds
the device will immediately start to recalibrate,
interrupting the standby and deactivating the
wake-up program .
Содержание BlueShadow Detector 40D
Страница 1: ...Detector 40D Instructions Document no V7647 HPLC HPLC...
Страница 71: ...Appendix Installation Qualification IQ...