5.7 GLP menu
In the
GLP menu, statistical data on important parameters of the pump
are displayed.
Navigate through GLP menu
1.
Use the horizontal keys to navigate through the submenu.
2.
If no buttons are pressed, the display returns to the main menu after a
short time.
3.
The menu contains no editable parameters.
5.8 Program menu
The menu
Program is used to list and create programs. The pump can
save up to 19 programs and up to 100 program lines with the following
functions:
Controlling the flow rate.
Controlling the solvent composition.
Sending control signals (Events).
Navigate through Program menu
1.
Open a program number of your choosing.
2.
Use the horizontal arrow keys to navigate through the two lines
of the display.
3.
To change a value, press and change the value by the numeric pad.
4.
Move the cursor to the double arrow on the lower left line to navi-
gate to the next line or create a new line.
5.8.1 Create program
The use of the display is demonstrated using the example of a program
to be created:
Program 01 is used to create a solvent gradient (channels A and B) with
constant flow of 100 ml/min. The percentage of solvent A is to be conti-
nuously decreased from 100 % to 0 % within 0.50 minutes.
Process
Figure
1.
Select the
Device program menu.
2.
Select the desired program number.
3.
The parameters for time 0.00 min can be chan-
ged by navigation with the horizontal
keys to the different parameters like
time/flow
1
, gradient composition
2
and
events
3
.
4.
Navigate to the double arrow on the lower left
side of the display and press the downward
key to create a new line.
5.
Enter the time value of the new line and adjust
the respective parameters.
6.
Press for two seconds to exit the submenu.
The program is automatically saved.
Procedure
Procedure
3