35
Operation
BlueShadow Pump 80P Instructions V7685
5.6.3 Device setting parameters
Menu
Explanation
Figure
Cntl. Interface
Setting for external communi-
cation connection of the pump.
The connections are on the rear
side of the device.
Serial: RS232-11520, RS232-
38400, RS232-19200, RS232-
9600 serial interface
LAN-DHCP: Automatic net-
work configuration
LAN manual: Manual net-
work-configuration (Port, IP
address, IP mask, Gateway)
Analog: Manual control of the
pump
Note:
After changing the
settings for external commu-
nication, the device should be
restarted.
Pr. limits
Settings for the minimum and
maximum pressure. The pump
stops automatically if the pres-
sure drops below the minimum
value or exceeds the maximum
value. The minimum pressure
value is used to ensure that the
pump does not run dry. If the
minimum is set to 0, the min-
imum system pressure is not
monitored.
Pump head
Pump head selection:
automatic detection
100 ml
250 ml
500 ml
1000 ml
Note:
The manual selection
of the pump head size is only
possible if the automatic recog-
nition of the pump could not
be performed. The selection is
used as a basis for calculating
the flow rate and determines
the maximum system pressure.
This value applies for all ope-
rating modes and cannot be
exceeded.