17
Installation
BlueShadow Pump 80P Instructions V7685
4.3.2 Connecting the eluent line
Make sure that all connections are tight.
Make sure that the tapered side of the cutting ring is pointed towards
the fastening screw of the Teflon tube.
Pump head
Process
Figure
100 ml
250 ml
500 ml
1.
Push the Teflon hose
1
through the fastening screw
2
and the cutting ring
3
.
2.
Insert the hose end as far as
possible into the eluent inlet
fitting
4
of the pump head.
3.
Tighten the fastening screw
by hand.
Fig. 8:
Connecting the eluent
line to the pump head
Pump head
Process
Figure
1000 ml
1.
Push the Teflon hose directly
on the olive-type tube fitting
1
.
2.
Fasten the Teflon hose with
hose clamp.
Fig. 9:
Connecting the eluent
line to the 1000 ml pump
head
Preparations
2
1
3
4
1