10
Product information
BlueShadow Pump 80P Instructions V7685
3.4 Pump heads
Pump head for use in preparative applications:
Standard model, stainless steel
Titanium inlays for biocompatible applications: 100 ml, 250 ml, 500 ml,
1000 ml
For biocompatibility, pump heads with titanium inlays can be used. The
pump heads can be exchanged by the user.
The front of the pump head is labeled with the maximum pumping capac-
ity and the pump head inlays (
SSt
for stainless steel,
Ti
for titanium).
Fig. 3:
Pump head 500 ml, stainless steel
The pump head is equipped with an RFID chip. It is used to monitor and
save all important parameters and settings of the pump and pump head.
The RFID technology offers the following advantages:
Automatic detection of all important pump head parameters by means
of radio signals (radio frequency identification), by the pump software
or the chromatography software:
—
Pump head type
—
Serial number and year of manufacture
—
Number of cycles and operating times
—
Limit values of the pump head parameters
All measuring data archived in accordance with GLP (Good Laboratory
Practice).
Transfer of all data to the software and pump display.
Designation
RFID detection