Application Examples
AZURA V 2.1S User Manual, V6860, Version 2.2
Result
The sample is injected and measured. The syringe remains connected until the
next injection.
Application Examples
This chapter describes possible applications for the different valve types. Of
course there are also other applications possible which are not listed here.
Valve Positions
The eluent flows through the sample loop in opposite direction of the position
load (see arrows in the figures).
1. The sample loop is filled by injecting the sample at port 1. The sample runs
into the sample loop (port 2 and 5) and excessive sample is transported into
the waste container through port 6.
2. The valve is switched from load to inject.
3. The pump (port 4) transports the sample to the column (port 3).
4. The eluent flows through the sample loop in opposite direction then during
injection, see arrows in the figures.
Valve Position
Load (L)
In the valve position
load
the sample is filled into the sample loop.
2. Fill the injection syringe
3
with
eluent. Make sure that there are no
air bubbles.
3. Insert the injection syringe into the
syringe port
2
.
4. Fill the sample loop
1
. Leave the
syringe connected.
Fig. 18 injection syringe
5. Press [
L
].
Fig. 19 display inject
Process
Figure
1
2
3