26
Installation
AZURA V 2.1S User Manual, V6860, Version 2.2
Prerequisite
There is a patch cable for the connection.
Process
1. Check that the IP address range of the router and of the company network
do not overlap.
2. In case of an overlap, change the IP address range of the router.
3. Use the patch cable to connect the router WAN port to the company net-
work.
4. Restart all devices, including the computer.
Controlling Several Systems Separately in a LAN
Devices connected to a LAN communicate through ports, which are part of the
IP address. If more than one HPLC system is connected to the same LAN and
you plan on controlling them separately, you can use different ports to avoid
interference. Therefore, the port number for each device must be changed and
this same number must be entered into the device configuration of the chroma-
tography software. We recommend to use the same port number for all devices
in the same system.
Note:
The port is set to 10001 at the factory. You must use the same numbers
in the device configuration of the chromatography software as in the device,
otherwise the connection fails.
Process
1. Find out port number and change it on the device.
2. Enter the port number in the chromatography software.
Result
The connection is established.
Connecting the Remote Connector
To control one device through another, you use the remote connector. The sin-
gle ports are used to exchange control signals. You find explanations on the
connector assignment at the end of the procedure.
To use remote control, you have to connect cables to the pin header. Finally
connect the pins of the pin header to the spring contact of the spring strip,
which is a connection to the board.
Prerequisites
The device has been switched off.
The power plug has been pulled.
Tools
depressor tool
Содержание Azura V 2.1S
Страница 1: ...Valve Drive V 2 1S User Manual V6860 HPLC...