Installation
19
AZURA V 2.1S User Manual, V6860, Version 2.2
Basic Functions of the DIP Switches
Combination Possibilities per Valve Type
Screwing the Adapter Plate onto the Valve Drive
Tools
Phillips screwdriver, size 3
1. Attach the adapter plate the right way around. The openings for the screws
point to the front.
2. Using the screwdriver, screw on the 2 screws (M3).
DIP
I/0
S6
S12
S16
1
ON
6P-multiposition
valve
12P-multiposition
valve
not permitted
OFF
6P/3C
not permitted
16P-multiposition
valve
2
ON
BIN 0-2 at 6P-multi-
position valves
BIN 0-3
BIN 0-3
OFF
RS-232, manual,
Pos. 1; [I]; [L]
RS-232, manual,
Pos. 1; [I]; [L]
RS-232, manual,
Pos. 1; [I]; [L]
3
ON
output OUT/BIN 3
output OUT/BIN 3
output OUT/BIN 3
4
OFF
3
OFF
not permitted
input OUT/BIN 3
input OUT/BIN 3
4
ON
Drive
Valve
Mode
DIP1
DIP2
DIP3
DIP4
V6
6P/3C
analog/RS-232
OFF
OFF
ON
OFF
6P-multiposi-
tion valve
analog/RS-232
ON
OFF
ON
OFF
6P-multiposi-
tion valve
binary
ON
ON
ON
OFF
V12
12P-multipo-
sition valve
analog/RS-232
ON
OFF
ON
OFF
12P-multipo-
sition valve
binary
ON
ON
OFF
ON
V16
16P-multipo-
sition valve
analog/RS-232
OFF
OFF
ON
OFF
16P-multipo-
sition valve
binary
OFF
ON
OFF
ON
Содержание Azura V 2.1S
Страница 1: ...Valve Drive V 2 1S User Manual V6860 HPLC...