KNAUER
P 6.1L User Manual V6890 Version 1.2
49
Maintenance and Care
Tool
Open-end wrench, size 10
Next steps
Choose a storage location according to the ambient conditions
listed in the respective chapter of this user manual.
Storage
Note that prior to storing all tubes and capillary should be emp-
tied or filled with an appropriate flushing solution (e.g. isopro-
panol). To prevent algae formation, do not use pure water. Seal
all inlets and outlets with cap fittings.
The device can be stored under the following ambient condi-
tions:
Disconnecting from Power Supply
Prerequisite
The device has been switched off.
Next steps
Disconnect all remaining electrical connections, remove the
pump head and pack the device for transport or storage.
Preparing the Pump Head for Storage
Prerequisites
The power plug has been pulled.
Auxiliary material
Syringe
Flushing solution:
CAUTION!
Damage to the pump head by residual chemicals.
If the pump head is stored and re-used later,
flush it and fill it with isopropanol. Afterwards,
seal the inlets and outlets with cap fittings.
Procedure
Process
1. Remove the solvent tubing and seal all open connectors
with cap fittings.
2. Remove the pump from the HPLC flow system and seal the
open connector of the pressure sensor (isocratic version) or
the mixer (binary or quaternary version) with cap fittings.
Temperature range
10–40 °C (50–104 °F)
Humidity
below 90 %, non-condensing
Procedure
Process
1. Pull the power plug from the socket and afterwards
remove from the device.
2. Pack the power cable with the device.