Maintenance and Care
Maintenance Intervals
Operating hours
The Control Unit, the Mobile Control and the different software
products (e.g. ClarityChrom
®
or OpenLAB
®
) enable you to dis-
play or read out the operating time of the pump. You find a
detailed description on how to display or read out GLP data in
the respective user manuals.
Cleaning and Caring for the Device
CAUTION!
Intruding liquids can cause damage to the device.
- Place solvent bottles next to the device or in a
solvent tray.
- Moisten the cleaning cloth only slightly.
All smooth surfaces of the device can be cleaned with a mild,
commercially available cleaning solution, or with isopropanol.
Checking the Fittings
Check if all fittings are tight. If fittings are not tight, re-tighten
fittings. Not the torque applicable for each fitting in order to not
damage any components.
Operating hours
Measures
1000
Check the torque of the screw fit-
tings
Clean the pistons of the pump
Check the check valves of the pump
head
5000
Replace all seals
Clean the check valves of the pump
head
10000
Replace pump head spare parts
Replace the check valves of the
pump head
Pump head inlet and outlet
Torque
Stainless steel fittings
7.5 Nm
PEEK fittings
3.5 Nm
Inline filter fittings
Torque
Stainless steel fittings
7.5 Nm
PEEK fittings
3.5 Nm
Capillary fittings
Torque
Stainless steel fittings
5 Nm