
83
Rear panel I/O
AZURA® Detector ECD 2.1 Instructions, V6685
11.6 Chassis grounding stud
On the rear panel a chassis grounding stud is available at the lower
right-side, next to the ventilation holes of the power supply compartment.
This grounding stud which is connected to the central grounding point of
the instrument can be used for shielding purposes. For example, to shield
the flow cell outlet from external electrical interferences (troubleshooting
noise issues), or to shield the flow cell from other equipment which might
be coupled in series with the electrochemical detector such as an MS.
Fig. 70
Left: ECD 2.1 rear panel grounding stud. Right: example of shielding
the flow cell by grounding the solvent outlet tubing.
Use the chassis grounding stud for shielding only, not for safety
grounding.
Содержание Azura ECD 2.1
Страница 1: ...HPLC Detector ECD 2 1 Instructions Document No V6685 ...
Страница 31: ...27 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 5 2 Overview of ECD 2 1 screens DC mode ...
Страница 32: ...28 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Pulse mode ...
Страница 33: ...29 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Scan mode CONFIC menu ...
Страница 34: ...30 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 DIAG menu ...