29
Operation
AZURA® Detector UVD 2.1S Instructions, V6820
Process
Figure
3.
Keep left arrow key pressed. Press right
arrow key to scroll through the IP ad-
dress, subnet mask, gateway and IP port
submenus.
4.
Let go of left arrow key.
5.
Using both arrow keys, edit the settings.
Fig. 16
Example for
submenu
5.4.4 Activating/deactivating the lamp
The status display shows not only absorption and wavelength, but also
indicates the status of the lamp.
HEA: Lamp is heating.
OFF: Lamp is deactivated.
ON: Lamp is activated.
Status display is active.
Process
Figure
1.
Keep left arrow key pressed. Press right ar-
row key until the correct display appears.
2.
Let go of left arrow key.
3.
Using both arrow keys, activate or deacti-
vate the deuterium lamp.
Fig. 17
Display lamp
When setting is finished, the status display appears.
5.4.5 Adjusting the chromatogram
To get a better representation of the chromatogram, you can either
smooth the output signal or adjust the voltage/absorption ratio.
Smoothing output signal
To smooth the output signal, you have to change the values of the time
constant.
0.00 s, 0.02 s, 0.05 s, 0.1 s, 0.2 s, 0.5 s, 1 s, 2 s
Status display is active.
Process
Figure
1.
Keep left arrow key pressed. Press right ar-
row key until the correct display appears.
2.
Let go of left arrow key.
3.
Using both arrow keys, set a value.
Fig. 18
Display time
constant
When setting is finished, the status display appears.
IP Addr1
192
Prerequisite
Procedure
LAMP D2
OFF
Result
Prerequisite
Procedure
TC:
0.5s
Result