31
Operation
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
The LEDs can have different colors depending on the operating
condi tions.
To start the standby, keep the standby button pressed for 5 seconds.
Note:
Malfunctioning system after repeated standby possible. After repe-
atedly using standby, restart the device using the power switch to reset
the device‘s data storage.
Color
Status
Operation
Left LED
red
Error
Check the system.
Shortly press the
switch to deacti-
vate the error
mes sage.
green
3D data are acquired.
Center LED
does not light
The lamp has been
switched off.
flashes green
The lamp/lamps are
initializing or the vali-
dation is progressing.
Wait until the
lamp is running or
the validation is
finis hed.
green
The deuterium lamp
is active.
Right LED
green
The device has been
switched on.
flashes green
The device not ready
for operation.
Wait until the
device is ready for
operation.
blue
The device is in
standby
Press the standby
button to end the
standby.
5.5 Default settings
Using the Mobile Control, you can reset the detector to its default
set tings.
Parameter
Setting
Network
LAN DHCP, port 10001
Lamps
Deuterium ON, Halogen OFF (DAD 6.1L only)
Time constant
2 s
Channels
Channel 1: WL = 254 nm, BW = 8 nm
Reference
correc tion
Reference channel enabled for channel 2,
WL = 360 nm, BW = 30 nm
Standby