27
Installation and initial startup
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
2
3
1
Fig. 20:
Depressor tool and pin header
Process
1.
Insert the depressor tool
3
in an upper small opening at the front
of the pin header
1
.
2.
Insert the cable into the opening
2
underneath the inserted de-
pressor tool
3.
Pull out the depressor tool.
Check whether the cables are tightly attached. Press the pin header onto
the connector. Finish the installation. Bring the device into operation.
4.8.9 Integrator connector
The integrator output supplies the current signal value as an analog vol-
tage (max. 5 V). The integrator output is connected to other instruments
with the analog connection cable or other special connection cables.
For details on the integrator connector, see the list below:
non-bipolar
4 channel
0 to 5 V
DAC 16 bit
scalable
Adjustable to offset
Next steps