AZURA® Assistant ASM 2.2L Instructions V6803
Product information 13
5.3.2
Rear view
Legend
1
Label with serial
numbers and article
numbers
2
CE-mark
3
Integrator output
4
Service interface
5
LAN-Connection 1
6
LAN-Connection 2
7
Pin header
8
Fan
9
Warning
bl
Mains connection
with mains switch
bm
On/off switch
bn
Warranty-Seal
Fig. 2
ASM 2.2L rear view
5.4
Device overview
The ASM 2.2L can be equipped with up to three modules.
These device types are available for selection:
Pump
Valve Drive
UV detector
In addition, the assistant can be operated with up to two empty modules.
5.5
Exceptions to the configuration
The following exceptions apply to the configuration
of the assistant:
1.
A maximum of one UV detector is allowed in an assistant.
2.
A maximum of two pumps is allowed in an assistant.
A high pressure gradient (HPG) is not supported.
3.
An assistant with two pumps and a detector AZURA® UVD 2.1S
is not supported because the power supply can be overloaded.
4.
An assistant with three empty modules is not supported.
1
2
4
3
56 7
9
bl
8
bm
bn
Device defect
The assistant must not be put into in operation with missing modules.
The ventilation system is thereby restricted.
Operate the assistant only fully equipped with modules
Insert empty modules if necessary.