AZURA® Assistant ASM 2.2L Instructions V6803
30
Initial startup
Connection
Function
EVENT 2 OUT
OC: TTL-compatible output
passive 5V (default setting with external pull-up to
24 V / 25 mA)
active 0 V
TTL output
passive 0 V
active 5 V
EVENT 1 OUT
OC: TTL-compatible output
passive 5V (default setting with external pull-up to
24 V / 25 mA)
active 0 V
TTL output
passive 0 V
active 5 V
START OUT
TTL-Ausgang
(default setting)/
OC with external pull-up to 24 V (25 mA)
Levels:
passive 5 V
active 0 V
The output can be used to start an external device.
START IN
TTL-compatible input
Low-active
Secure switching threshold at least 10 mA
The start of a module can be triggered by a signal
(short circuit to GND) from an external device.
ERROR OUT
TTL-output (
default setting) /
OC with external pull-up to 24 V (25 mA)
Levels:
passive 5 V
active 0 V
Output remains active until the cause of the error has
been eliminated.
ERROR IN
TTL-compatible input
Low-active
Secure switching threshold at least 10 mA
For a signal (short circuit to GND) from a
external device an error message appears and
the device stops.
AZ IN
TTL-compatible input
Low-active
Secure switching threshold at least 10 mA
For a signal (short circuit to GND) from a
external device, an autozero is performed by the
detector.