AZURA® Assistant ASM 2.2L Instructions V6803
Initial startup 29
Procedure
Process
Figure
1.
Insert the operating tool
3
into an upper small opening on
the front of the terminal strip
1
.
2.
Lead the cable into the open-
ing
2
below the inserted
operating tool.
3.
Pull out the depressor tool.
Fig. 13
Pin header
Check if the cables are firmly attached.
Press the pin header onto the connector.
Finish the installation.
Bring the device into operation.
8.3.2
Pin header assignments
Fig. 14
Pin header assignments
Connection
Function
GND
Reference point of the voltage at the signal
inputs.
ANA OUT
Voltage range 0 V – 10 V
Depending on the module in the ASM 2.2L, the pump
pressure or the detector signal can be output.
ANA IN
Voltage range 0 V – 10 V
Depending on the module in the ASM 2.2L, the flow
rate or the wavelength can be set.
1
2
3
Next steps
1 2 3 4 5 6 7 8 9 10 11 12
GND
ANA
OUT
ANA
IN
EVENT
2 OUT
EVENT
1 OUT
ST
AR
T OUT
ST
AR
T IN
ERROR OUT
ERROR IN
AZ IN
+4-20 MA
GND