AZURA® Assistant ASM 2.2L Instructions V6803
22
Initial startup
Device defect
Changes of the environmental temperature cause condensation inside
the device.
Allow device to acclimate for 3 h before connecting to power supply
and taking into operation.
8.1 Connecting the leak management
The leak management consists of the leak sensor and the drainage system.
The drainage system ensures that escaping liquids flow into a waste bottle.
If there is too much liquid, the red LED starts flashing. Both the device and
the data acquisition via chromatography software are stopped.
Remove the front cover.
Procedure
Process
Figure
1. Carefully push the funnel
1
into the center opening of the
capillary guide
2
.
1
2
Fig. 8
Funnel and capillary guide
2. Push the long ending of the first
nozzle
4
into the hose
3
.
3
4
Fig. 9
Hose and nozzle
3. Afterwards, push the nozzle
onto the funnel.
4. Push the other end of the hose
onto the nozzle
5
of the leak
tray.
5
Fig. 10
Connect the hose to the
device
Prerequisites