Troubleshooting

Further measures

Inform the Technical Support of KNAUER.

System messages

If other system messages are displayed besides those listed below, please 
turn the device off and then on. Inform the Technical Support of KNAUER 
in case the system message repeats itself.
The system messages are sorted alphabetically. 

Degasser

Air bubbles occur in the 
solvent tubes of the 
degasser outlet.

Check the inlet and outlet fittings. Replace 
old fittings, if necessary.

No eluent flow

If a buffer solution stays in the degasser for 
an extended period, it can cause clogging.
1. Use another degasser chamber. 
2. Carefully let water flow into the clogged 

degasser chamber to dissolve the buf-
fer. 

3. If this does not help, contact the Techni-

cal support.

SmartMix 

mixer

The pump was switched 
off automatically, because 
Pmax was exceeded.

Exchange the filter system.

Pressure increase

Replace cartridge.

Filter system inserted 
wrong

Exchange the filter system.

Problem

Solution

System message

Solution

A

At least one wavelength 
must be valid.

Check whether a channel is on. Check 
whether the wavelengths are within per-
missible range (190–500 nm).

C

Calibration failed

Switch the device off and on. Check 
whether lamps, motor and filter are func-
tioning correctly. Inform the Technical Sup-
port of KNAUER in case the system 
message repeats itself. Restart calibration 
on the device or in the chromatography 
software.

Cannot initialize

Check cables and connections in local area 
network.

Cannot proceed: D2 
lamp heating

Wait for the ignition of the D2 lamp.

Cannot proceed: D2 
lamp is OFF!

Switch D2 lamp on before "RUN".

Cannot proceed: Lamps 
are off

Switch the lamps on the device to "ON".

Cannot read RTC

Switch the device off and on. Inform the 
Technical Support of KNAUER in case the 
system message repeats itself.

Содержание Azura Assistant ASM 2.1L

Страница 1: ...Assistant ASM 2 1L User Manual HPLC...

Страница 2: ...AUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 D 14163 Berlin Phone 49 30 809727 0 Fax 49 30 8015010 Internet www knauer net E Mail info knauer net Version information Article number V6815 Version...

Страница 3: ...Safety for users 11 Signal words 12 Decontamination 12 Symbols and signs 12 Unpacking and setup 13 Contacting the Technical Support 13 Location requirements 13 Unpacking 13 Connectors on the rear sid...

Страница 4: ...ce contract 27 Required system information 27 Cleaning and caring for the device 27 Maintenance tasks for users 27 Technical data 28 Communication 28 General 28 Repeat orders 28 Legal information 30 T...

Страница 5: ...s Environmental analyses Eluents Small quantities of substances such as additives modifiers or salts can influence the durability of the materials For further reference on chemical compatibility pleas...

Страница 6: ...take a look into the UVD 2 1S user manual V6820 http www knauer net fileadmin user_u pload produkte files Dokumente detectors azura manuals V6820_A ZURA_UVD_2 1S_User Manual_EN pdf Flow cells Flow cel...

Страница 7: ...h 6 ports and 3 channels In case of multiposition valves the central port does not count For further information take a look into the V 2 1S user manual V6860 http www knauer net fileadmin user_upload...

Страница 8: ...ood or similar extraction device KNAUER Valves Ports Connection Material Pressure stability 2 position valve 6 1 16 1 8 Stainless steel 300 bar 2 position valve 6 1 16 Stainless steel 400 bar 2 positi...

Страница 9: ...ight of detector UVD 2 1S because the fittings of the detector flow cell cannot be mounted An assistant with only degasser integrated is not allowed Position of modules and IP port connections The pos...

Страница 10: ...Devices ClarityChrom OpenLab Mobile Control V4 Two pumps independent yes no no Fraction valve 1 1 no Injection module1 1 An injection module is a combination of one pump and one 6 port 2 position valv...

Страница 11: ...ts tools and solvents made or recommended by KNAUER Good Laboratory Practice GLP Accident prevention regulations published by the accident insurance companies for laboratory work Filtration of substan...

Страница 12: ...occur Minor injuries can occur Device defects can occur Decontamination Contamination of devices with toxic infectious or radioactive substances poses a hazard for all persons during operation repair...

Страница 13: ...t 10 cm if further devices are set up on both sides At least 15 cm on the rear panel for the fan Leave the power plug on the rear of the device accessible to be able to disconnect the device from the...

Страница 14: ...ou notice any damage contact the Technical Support 6 Place the device in its site of operation and remove protective foil Next steps Store packaging and keep the included packing list for repeat order...

Страница 15: ...t least 1000 ms START IN TTL input Low active Secure switching threshold at least 1000 ms After receiving a signal short circuit to ground from an external device the device starts If controlled with...

Страница 16: ...w on page 5 PEEK fittings Component defect Damage to the ports caused by strongly tightened fittings Use 5 Nm torque for stainless steel fittings Use 0 5 Nm torque for PEEK fittings ASM 2 1L as Compac...

Страница 17: ...nt cover Process Process Figure 1 Carefully push the funnel 1 into the center opening of the capillary guide 2 Fig 7 Funnel and capillary guide 2 Push the long ending of the first nozzle 4 into the ho...

Страница 18: ...their IP address Prerequisite In Windows power saving hibernation standby and screen saver must be deactived In case you use an USB to COM box the option Allow the computer to turn off ths device to...

Страница 19: ...e can be used to remotely control the system Integrating the LAN into a Company Network A network administrator can integrate the LAN into your company net work In this case you use the WAN port of th...

Страница 20: ...trol is a program for a computer or tablet To operate via Mobile Control the computer or tablet needs to be connected to a WLAN router You will find a detailed description in the Mobile Control user m...

Страница 21: ...odules are at the desired location The power plugs of the modules are plugged in The LAN connections between the modules and router are connected The LAN cable is connected to the workstation and rout...

Страница 22: ...ation Qualifica tion is a standardized document and free of charge It is not part of the delivery please contact the Technical Support in case of request The Operation Qualification includes the follo...

Страница 23: ...r is integrated into a company network pull out the patch cable from the WAN port Can the devices communicate with the computer even though the router is disconnected from the company network 5 Turn o...

Страница 24: ...ves When the pump head seals are defec tive solvent enters the piston backflush ing Inform the Technical Support of KNAUER Exchange the pump head Pressure and flow rate variations Pump without pressur...

Страница 25: ...ed offautomatically because Pmax was exceeded Exchange the filter system Pressure increase Replace cartridge Filter system inserted wrong Exchange the filter system Problem Solution System message Sol...

Страница 26: ...t Switch the device off and then on If the leak sensor is still not present contact the Technical Support of KNAUER Leak was detected Switch off the device Remove the leak and start the device afterwa...

Страница 27: ...endar week In the example the module was manufactured in calendar week 32 of the year 2010 The last five digits are the identification number for the interface card Interface Serial number Cleaning an...

Страница 28: ...ts and accessories can be found online www knauer net Module Measures 2000 operating hours Detector Check light intensity of the D2 lamp Inspect the flow cell assembly Clean the flow cell Replace the...

Страница 29: ...ec tion valve pH flow cell and 2 small prepacked col umns A9854 1 Mounting bracket AZURA L for axial compressible columns with 20 mm ID A9853 3 Mounting bracket AZURA L for Hypershear mixing chamber A...

Страница 30: ...nd transport damage In the event of device malfunctions directly contact the manufacturer KNAUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 14163 Berlin Germany Phone 49 30 809727 111 Telefax 49 30...

Страница 31: ...uld be as small as possible Degasser Degasser module for fluids GLP Good Laboratory Practice quality assurance for labo ratories Gradient Time dependent composition of solvent mobile phase on low pres...

Страница 32: ...ASM 2 1L user manual V6815 Abbreviations and terminology 32 Solvents Mobile phase eluent or carrier for liquid chromato graphy Term Explanations...

Страница 33: ...M 2 1L 29 DG 2 1S 8 P 2 1S P 4 1S 6 UVD 2 1S 6 V 2 1S 7 Mobile Control 20 29 mounting bracket 29 O operating environment 13 OQ 22 P packing list 14 PEEK 16 port LAN 19 power disconnect 13 plug 13 powe...

Страница 34: ...ASM 2 1L user manual V6815 Index 34 V valve 7 W warranty 30...

Страница 35: ...R manuals online www knauer net knowledge KNAUER Wissenschaftliche Ger te GmbH Hegauer Weg 38 14163 Berlin Phone Fax E Mail Internet 49 30 809727 0 49 30 8015010 info knauer net www knauer net KNAUER...

Отзывы: