41
Operation
AZURA® Pump P 2.1L Instructions V6840
Process
Figure
Procedure
1.
Switch on the pump at the
power switch
1
on the rear
side.
2.
Wait until the pump has com-
pleted the self-test.
3.
If the self-test has been suc-
cessfully completed, the LED
on the right lights up green.
Fig. 34:
1
Power switch
The pump is now ready for operation. If the test fails, an error message
will be displayed. Contact the Technical Support of the manufacturer if
the error occurs several times.
Forestall pressure fluctuations by letting the pump run for 1 hour.
5.4
Software control
There are several options for controlling the device:
with chromatography software
with KNAUER Mobile Control
Note:
It is not possible to use two control methods simultaneously. If the
device is connected to the software, it cannot be controlled via Mobile
Control.
5.4.1
Controlling with chromatography software
To operate the device with software, you have to establish a connection
between the LAN port and a computer. Devices can be controlled with
e. g. OpenLAB EZChrom edition version A.04.05 or higher, ChromGate
version 3.3.2 or higher and ClarityChrom version 3.0.7 or higher.
You find a detailed description on chromatography software in the corre-
sponding instructions.
5.4.2
Controlling with Mobile Control
The Mobile Control is a software which can be installed on your compu-
ter or tablet. To control the device using the Mobile Control, connect the
com puter or tablet with operating system Windows 10 to a wireless LAN
router. The firmware version of the pump must be V01.09 or higher. You
find a detailed description in the Mobile Control Instructions (V6851-2).
Result
Next step