5.2
Purging the pump
Before the pump can be used, it must be purged to remove excess air in
the capillaries.
Purge the pump in the following cases:
At initial startup to eliminate air bubbles in hoses and capillaries.
When changing solvents.
After using buffer solutions to eliminate salt residues.
Before turning off, if you do not plan to restart the device within
shortly.
The installation has been completed.
The capillaries and hoses have been connected.
The pump has been switched on and is in „flow mode“.
Syringe with Luer lock
Use the purging solvent to be used in the following application.
Note:
The purging process may take a while during initial operation
because the solvent tubes are filled with liquid for the first time.
Note:
If a buffer solution is used, select a solvent for purging in which the
buffer solution is soluble.
Column defect
Damage to the column due to purging.
Open the venting screw.
Remove the column.
When is flushing
required?
Prerequisites
Tool