AZURA® Pump P 2.1L Instructions V6840
36
Installation and initial startup
4.11.2 Connecting the pin header
To control one device through another, the pin header is used.
To use remote control, you have to connect cables to the pin header.
The single ports are used to exchange control signals.
The device is switched off.
The power plug is disconnected from the device.
Depressor tool
Electronic defect
Connecting cables to the multi-pin connector of a switched on device
cau ses a short circuit.
Turn off the device before connecting cables.
Pull the power plug.
Electronic defect
Electrostatic discharge can destroy the electronics.
Wear a protective bracelet against electrostatic discharge and ground.
2
3
1
Fig. 32:
Depressor tool
Process
1.
Insert the depressor tool
3
in an upper small opening at the front of
the pin header
1
.
2.
Insert the cable into the opening
2
underneath the inserted
depressor tool
3.
Pull out the depressor tool.
Check whether the cables are tightly attached. Press the pin header onto
the connector. Finish the installation. Bring the device into operation.
Prerequisites
Tools
Procedure
Next steps