Installation and initial startup
AZURA® Detector UVD 2.1L instructions, V6831
2
3
1
Fig. 18:
Operating tool
Process
1.
Insert the operating tool
3
in an upper small opening at the front of
the pin header
1
.
2.
Insert the cable into the opening
2
underneath the inserted opera-
ting tool.
3.
Pull out the operating tool.
Check whether the cables are tightly attached.
Press the pin header onto the connector.
Finish the installation.
Bring the device into operation.
4.11 Switching on
The device is switched on by the power key on the back of the device.
Installation has been completed.
Flow cell is mounted on the detector.
Steps
Figure
1.
Insert the plug into the
female connector
2
at the rear panel.
Fig. 19
Power switch with female connector
2.
Plug the other end into
an electric socket.
3.
Switch on the device at
the power switch
1
.
The detector starts its self calibration. The end of the self calibration is in-
dicated by the flashing of the status LED. If the self calibration fails, switch
the detector off and on again.
Next steps
Prerequisite
Procedure
1
2
Result