Installation and initial startup
AZURA® Detector UVD 2.1L instructions, V6831
4.10.2 Analog control
Using the analog port, you can control the wavelength by changing the
applied voltage. A Mobile Control is required in order to select the option
ANALOG in the SETUP menu.
To use the analog port for controlling the detector, you have to set a zero
point and enter a scaling value.
Zero point at 0 V = 000 nm
Scaling: 100 nm per Volt
If 5 V voltage is applied, the wavelength is 500 nm.
4.10.3 Integrator connector
The integrator connector sends measuring signals from the detector.
non-bipolar
1 channel
0 to 5 V
DAC 20 bit
scalable
adjustable to offset
4.10.4 Connecting the pin header
To control one device through another, the pin header is used. To use
remote control, you have to connect cables to the pin header. The single
ports are used to exchange control signals.
The device is switched off.
The power plug is disconnected from the device.
Operating tool
Electronic defect
Connecting cables to the multi-pin connector of a switched on device
cau ses a short circuit.
Turn off the device before connecting cables.
Pull the power plug.
Electronic defect
Electrostatic discharge can destroy the electronics.
Wear a protective bracelet against electrostatic discharge and ground.