24
Installation and initial startup
AZURA® Detector UVD 2.1L instructions, V6831
4.10 Remote control
For receiving start, control, and error signals from external devices
For sending start, control and error signals to external devices
Fig. 17:
Relation between display and Events terminal strip
4.10.1
Connector assignment
Plug
Function
EV 1
(Event 1)
Relay contact
The contact is on a floating basis. Its setting de-
pends on the settings in the software.
Steady-rate signal:
- passive = open relay contact
- active = closed relay contact
Pulse:
- closed relay contact for at least 1000 ms
- permissible load of the relay contact: 1 A/24 V DC
EV 2
(Event 2)
TTL compatible output
Levels:
- passive 5 V
- active 0 V
Pulse:
- 0 V for at least 1000 ms
EV 3
(Event 3)
TTL compatible output
Levels:
- passive 5 V
- active 0 V
Pulse:
- 0 V for at least 1000 ms