Revision 5.5
91
8
Maintenance
8.1
Periodic Maintenance
NOTICE
Repairs may be performed solely by authorized service technicians of Gebrüder Martin.
Inspection/Maintenance
Maintenance interval
To be performed by
Control of the
EMERGENCY STOP
switch
prior to the first surgery
on each day of use
Hospital staff
Inspection of accessory equipment
prior to each surgery
Hospital staff
Inspection of the pilot beam for coaxial
exit from the focusing handpiece
prior to each surgery
Hospital staff
Verification that CO
2
beam and pilot
laser beam match
prior to each surgery
Hospital staff
Control of the laser warning lamp (s)
prior to each surgery
Hospital staff
Cleaning and disinfecting the unit's
external surfaces
after each surgery
Hospital staff
Inspection of the electrical connections
and cables for potential defects
every 6 months
Hospital technician
Inspection/replacement of the air
cooler filter
1 × annually
(during safety checks)
Martin Service Center
Inspection of the cooling water level or
refilling of cooling water, respectively
1 × annually
(during safety checks)
Martin Service Center
Inspection of the output power level
1 × annually
(during safety checks)
Martin Service Center
Replacement of the ultrapure air filter
every 2 years
(during safety checks)
Martin Service Center
Replacement of cooling water
0.75 l distilled water and
0.75 l propylene glycol
every 2 years
(during safety checks)
Martin Service Center
Safety Checks (SC)
required once per year
Martin Service Center