![KLS Martin MCO 25 plus Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/kls-martin/mco-25-plus/mco-25-plus_instructions-for-use-manual_1979006056.webp)
56
Revision 5.5
6.5.4
Display Fields on the Display
Observe Fig. 6-3, page 52. and Fig. 6-4 page 53.
No. Mode
Name
Function
Description
D1
F, C
Power,
Max Power
Laser output power in W, maximum
power in pulse or multiple pulse mode
2–25 W cont. (F1),
10–25 W pulse and
scan (F2, F3, C1…C5)
D2
F
Pulse duration
Duration of the pulse in ms or s
5 ms–10 s (F2, F3)
D2
C
Shape size
Side length or diameter of the scanned
shape in mm
See table for sizes
D3
TM
Pause
Time between 2 pulses,
displayed in ms or s
5 ms–10 s (F3)
5 ms–0.1 s (F4)
D3
FM
Pulse rate
Pulse frequency or pulse rate, number
of pulses per second in Hz
0.1–5 Hz (F3)
0.1–200 Hz (F4)
D3
F4
Average super
pulse output
Super pulse output
0.3–11 W
D3
C
Scanner
scanning speed
Selects the speed of the scanned laser
light
slow (1-fold)
medium (2-fold)
fast (4-fold)
D3
SC
Scanner
overlap
Frequency of the super pulse rate
light (0 %), medium
(50 %),
strong (80 %) overlapping
D3
FC
Scanner
fill factor
Amount of treated area
light (20 %), medium
(50 %), strong (80 %)
D4
Cycle length
Duration of a cycle in ms or s
0.1–10 s (F3, F4)
D4
C
Handpiece
Selects the scanner handpiece
Scanner handpiece 127
and 200 mm,
micromanipulators and
special optics
D5
TM
Cycle pause
Time between 2 cycles,
displayed in ms or s
0.1–10 s (F3, F4)
10 ms–10 s (C)
D5
FM
Cycle rate
Cycle interval frequency or cycle rate,
number of cycles per second in Hz
0.1–5 Hz (F3, F4, C)
D6
Pilot
Intensity of pilot laser light,
operating mode
Adjustable brightness
1–100%
D7
F
Average output
Average laser energy calculated from
pulse duration, pause time and
maximum power
0.1–25 W, is calculated
and displayed
D7
C
Energy area
density
Laser energy per area (constant when
changing shape/size)
Energy area density in
J/cm². Calculated from
result of scanning speed,
handpiece and output.
D8
Displays
function symbol
Laser mode symbol,
scan shape with type and orientation
Function key symbol and
selection for scanner
function
D8
C5
Line scan shape Shape of line with rotation
Rotation of line scan
shape with keys P2 and P3
Tab. 6-3:
TM = Time Mode, FM = Frequency Mode
(F) = Function F1…F5, (C) = Scanner function C1…C5,
(SC) = Super pulse scanning, (FC) = Fractionated scanning