5.8
Preparing the Mirror Joint Arm for Use with a Scanner
In order to compensate the weight of the scanner handpiece, which is higher than that of other
focusing handpieces, the tension of the mirror joint arm must be set to the stronger value.
•
Unburden the mirror joint arm by pulling it towards the user.
•
Change arm latch (1) from
locked
position to
unlocked
position.
•
Swivel arm slightly backwards until adjusting lever (2) can be turned to the
scanner
position. This requires little effort if the arm has been sufficiently relaxed. If required, relax
arm further.
•
Now pull arm again forwards against the spring force and set arm latch (1) to
locked
position.
—
The arm now has the correct spring force for use with a scanner.