Revision 5.5
5.5.5
Foot Switch Connection
The foot switch cable (8) is to be connected to the 5-pin connection socket (6) and secured by
screwing on. The foot switch (9) must be positioned in a way that it is easily accessible for the
operator.
CAUTION
Risk of injury by accidental laser activation!
The foot switch is equipped with magnetic reed contacts. Magnetic fields in the immediate
vicinity can cause accidental activation of the foot switch and thus uncontrolled emission of
laser light.
•
Keep the foot switch away from magnetic fields (MRI in particular)!
•
Keep magnetic materials away from the foot switch!
6 Connection socket for foot switch
9 Foot switch
8 Connector on foot switch cable
10 Step-on protector
220-240V
~ 2 x T 4 AH, 250 V