Revision 5.5
31
5.5.2
Mains Connection
NOTICE
Use only the supplied mains cable or a replacement cable by Gebrüder Martin with the laser
system!
•
Plug mains cable into mains connection socket (1, Fig. 5-1).
The supply parameters are 220 –240 VAC, 50/60 Hz, max. 4 A.
•
Fuse the earthed mains connection socket for at least 6 A.
•
The supplied mains cable is a 3-wire cable (including protective ground wire).
NOTICE
Do not use splitting boxes or extension cords!
5.5.3
Equipotential Grounding Pin
The device can be connected to the potential equalization
system in the operating theater at the equipotential
grounding pin (5, Fig. 5-1) via the potential equalization
cable.
5.5.4
Mains Fuses
NOTICE
Fuses could be only replaced by service technicians of Gebrüder Martin or authorized service
partners.
After opening the cover (2, Fig. 5-1) you will have access to 2 micro-fuses whose electrical
parameters are printed on the retainer (T 4 AH, 250 V, slow-blow).
If current consumption is too high in the laser circuit, fuses protect the system.