maxium® smart C Electrosurgery Unit with maxium® smart Beam
Revision 1
55
10.3.3
Setting the Effect
The performance of the maxium® smart C can be changed via setting the effect. For this purpose, there are
several effect settings for all current types, which result in different degrees of hemostasis, regardless of
the selected power. When the effect is being adjusted, the corresponding voltage values and the change in
hemostasis are displayed in a diagram.
After selecting the effect to be adjusted by pressing,
a red frame appears around the effect value, and a
window is opened.
•
The effect can now be adjusted using the slider
and/or the
+
/
-
buttons.
Any adjustment can be confirmed by tapping the
check mark or canceled by tapping the cross.
10.3.4
Selecting the Activation Type/Foot Switch
The maxium® smart C can be activated via finger switches on the instrument, via foot switches, and for
bipolar coagulation currents also automatically. While each finger switch is always unambiguously assigned
to the channel to which its instrument is connected, foot switches must first be assigned to a channel by
the user.
Two foot switches can be connected to the maxium® smart C, a double-pedal foot switch with a yellow
pedal for cutting and a blue pedal for coagulation, as well as a single-pedal foot switch with a neutral black
pedal.
After pressing of the area on the right to the effect,
an activation menu is opened. The respective
activation type can be selected here.
The following selection options for activation are
available:
Monopolar cutting:
•
using the finger switch
•
using the yellow pedal
•
using the black pedal
Monopolar coagulation:
•
using the finger switch
•
using the blue pedal
•
using the black pedal