Installation of the maxium® smart C and the
maxium® smart Beam
8.1
Standalone Installation of the maxium® smart C
The maxium® smart C is designed to be set up on a level surface such as a table or a shelf.
Connection to a power outlet is established with the supplied mains cable to be inserted into the mains
port on the backside. The maxium® smart C can use any AC mains voltage in the nominal range of
100–240 V. Thus no special adaptation to the mains voltage at the installation is required.
If the maxium® smart C is to be connected on-site to a grounded equipotential bonding network, then this
can be done using the connection pin on the rear via the equipotential bonding cable, REF 80-260-50-04.
This cable is optional and not included in the scope of delivery of the maxium® smart C.
Usually, on the rear also a double-pedal switch will be connected. If required, in addition a small single-
pedal switch can be connected.
After connection of the electrode cables, the maxium® smart C is then ready for commissioning.
Risk of electric shock!
In order to avoid the risk of electric shock, this unit may be connected only to a grounded mains supply.
Risk of damage to property
In order to be able to completely separate the unit from the grid in the event of danger, either the power
socket of the unit or the outlet into which the mains cable is plugged in should remain accessible.
For the decommissioning of the unit, no special measures are necessary.