maxium® smart C Electrosurgery Unit with maxium® smart Beam
Revision 1
41
The gas supply at the electrode handle is part of a Luer-lock
connection and to be connected to the socket (1, Fig. 7-1) on the
front of the maxium® smart Beam. In order to flush out any air
present in the applicator, by pressing the
PURGE
key (2, Fig. 7-1) an
argon gas flow can be initiated without activating the
electrosurgery unit.
The components of the gas supply and the gas flow control inside the unit up to the outlet are non-sterile.
For more protection against microbial contamination, there is the possibility of inserting a sterile single use
filter (REF 80-181-90-04) with a pore width of 0.2 µm between the gas outlet of the unit and the sterilizable
applicator, which will hold back potential germs (with the exception of viruses).
7.2
Instruments, Electrode Handles, Electrodes, Foot Switches and Cables for
maxium® and maxium® smart Beam
The maxium® smart C as well as the maxium® smart Beam may be used only with accessories, wear parts
and single use articles whose safety of use has been demonstrated by a declaration of conformity. The use
of uncertified accessories by other manufacturers may constitute a source of danger. In case of doubt,
consult the manufacturer.
The accessories certified for electrosurgery units by Gebrüder Martin can be taken from the accessories
catalog of Gebrüder Martin, which can be requested or directly downloaded from www.klsmartin.com.
Combinations of accessories which are not mentioned in these Instructions for Use may be used only if
expressly designed for the intended application. Features and safety requirements must not be affected
adversely.