maxium® smart C Electrosurgery Unit with maxium® smart Beam
Revision 1
113
13.6
System Errors
Below, a number of messages are listed which the user can take care of on his own if necessary.
Message
Description
Overtemperature: Power output
interrupted!
The HF generator is overheated from high power output over a longer
period and needs a few minutes to cool down. As soon as the text
message disappears, re-activation is possible.
If this message is displayed after moderate power output a well, the
unit should be checked.
B1: Unknown bipol instrument
connected! Code: xx
B2: Unknown bipol instrument
connected! Code: xx
The bipolar instrument detection reads an unknown instrument code
and cannot support this instrument.
It may be a new instrument which is not known to the installed
software yet. In this case a software update will be necessary. If it is
an instrument which was always detected before, then connect the
instrument to the other Bipolar Output. If it is detected correctly
there, then there is a malfunction in the unit, otherwise the
instrument itself is defective.
Error 285: Board scan failed – please
restart system!
There was an error in the detection of the hardware configuration
during self-test. Turn the unit off and then on again. If the message
appears again, then there is a malfunction in the unit.
Error 298: Unit massively too cold, please
switch off for 1 h!
The unit is or was located in a cold environment. If is set up in a much
warmer environment, then there is the risk of condensation inside the
unit and thus an electrical safety risk. The unit may be put back into
operation only after the condensate has been completely evaporated.