![KLS Martin marLED X Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/kls-martin/marled-x/marled-x_service-manual_1978990064.webp)
64
Risk of property damage from electrostatic discharges (ESD)!
Electronic components in the light can be destroyed by ESD!
•
Work on the opened light body may be carried out only with sufficient ESD protection.
8.3
Exchange Works
8.3.1
Opening the Light Housing
Risk of property damage due to falling parts!
Never open the light as long as it is attached to the suspension arm system! To detach the light first, see
section 8.1.2 “Removing the Light Head from the Spring Arm”, page 60.
Contamination of the light!
Screws are secured by factory
default with a blue PA coating.
•
Make sure that no coating
residue remains inside the light
upon removal of the screws.
•
Remove the screws (1).
•
In case of a light wired with 9
pins, disconnect the 3 video
cables from the sockets (2).