4.2.4
Connections and Fuses
Fig. 4-1:
Connection designations and fuses
ASPS area only in case of integrated ASPS switching, N fused with F2 is used only for IT grids
The power module can be supplied with or without ASPS switching. Standard wiring is carried out via
N
and
L
, protected by
F1
in the inside of the power module.
In case of IT grids, the connection
N fused
, protected by
F2
, is used instead of
N
.
In case of the variant for ASPS, the battery input is protected by
F4
in the inside of the power module.
If the power module is supplied with mains voltage, the LED
AC ok
(1) that is visible from top through the
recesses in the housing is lit.