Gebrüder Martin GmbH & Co. KG
KLS Martin Platz 1
D-78532 Tuttlingen/Germany
5. Checking the monitor suspension
OK Not OK N/A
a) Checking and securing the bearing assembly (marView only)
☐
☐
☐
b) Checking the monitor bay (marView and TFT-FS)
☐
☐
☐
c) Checking the TFT bearing neck on the spring arm (marView and TFT-FS)
☐
☐
☐
d) Checking the TFT Duo cross-arm connection to the TFT individual suspension
(marView and TFT-FS)
☐
☐
☐
e) Checking the connection of the TFT Duo cross-arm to the spring arm
☐
☐
☐
6. Checking the light functions
OK Not OK N/A
a) Checking the emergency power switch
☐
☐
☐
b) Function test according to Instructions for Use
☐
☐
☐
c) Checking the light field
☐
☐
☐
d) Checking COM interface (e.g. querying of current firmware version)
☐
☐
☐
Firmware version
e) Functional check of the COM interface (e.g. illum/-)
☐
☐
☐
f) Functional check of the COM interface (e.g. command “camera zoom in/out”)
☐
☐
☐
g) Checking the illuminance
Enter specified value from Technical Data:
Specified:
Measured:
MP1:
MP2:
MP3:
MP4:
h) Checking the supply voltage on contact block Specified:
Measured:
MP1:
MP2:
MP3:
MP4:
at no load:
max. 36 VDC
at nominal load:
min. 24 VDC
at marLED® E3:
100–240 VAC