![KLS Martin diomax Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/kls-martin/diomax/diomax_instructions-for-use-manual_1978994066.webp)
Laser Surgical Unit diomax
®
66
Revision 6
10
Accessories
Accessories and disposable articles may be used only if they are CE-labeled and correspond to
the articles in the following list. The list of accessories may be updated and changed. The most
current list can be requested from the manufacturer. For cleaning, maintenance and
sterilization purposes, follow the instructions for use of the respective accessory equipment!
Accessory
Data
Item no.
Laser protective goggles, universal
DIR 980 L5
79-100-55-04
Focusing handpiece
3 attachments with
beam diameters of 0.5 / 1 / 1.5 mm
79-600-00-04
Bare Fiber, 300 µm
Laser fiber 300 µm × 3 m (5 pieces)
79-700-30-04
Bare Fiber, 400 µm
Laser fiber 400 µm × 3 m (5 pieces)
79-700-40-04
Bare Fiber, 600 µm
Laser fiber 600 µm × 3 m (5 pieces)
79-700-60-04
10.1
Safety Distances (NOHD)
For accessories there are the following safety distances (NOHD = Nominal Ocular Hazard
Distance) defined for which the irradiation strength or irradiation is under the approved limits.
Accessory
NOHD
Divergence
Data
Item no.
Focusing handpiece
9.0 m
100 mrad
-
79-600-00-04
Bare Fiber 300 µm
4.0 m
200 mrad
300 µm
79-700-30-04
Bare Fiber 400 µm
4.0 m
200 mrad
400 µm
79-700-40-04
Bare Fiber 600 µm
4.0 m
200 mrad
600 µm
79-700-60-04