![KLS Martin diomax Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/kls-martin/diomax/diomax_instructions-for-use-manual_1978994043.webp)
Laser Surgical Unit diomax
®
Revision 6
43
5.10
Switching off the Device
•
Turn key in the key switch from position
1
to
0
.
CAUTION
Risk of injury
Any laser device that is not being used must be protected against unauthorized use.
•
Remove the key from the key switch and store it in a safe location.
•
Remove laser fiber and handle according to instructions for use.
•
Clean the accessory equipment and store in a safe location.
•
Clean and disinfect the outer surfaces of the device with a cloth according to the hygiene
regulations. Make sure that no liquids can enter the device openings. Sprays must not be
used under any circumstances!
NOTICE
Moisture must not enter the SMA-X fiber socket.
5.11
Emergency Stop
•
Press the red EMERGENCY-STOP switch.
—
The laser immediately stops the emission of light. The device is fully disconnected from
the grid.
•
Turn key switch to position
0
.
•
Bring the
EMERGENCY-STOP switch
into the pulled position before reactivating the
device.
NOTICE
Do not use the EMERGENCY-STOP switch for regular switching off of the device!