Instruction
Manual
For Professionals...Since 1857
®
ENGLISH
MM6000
• TRUE RMS
• BACKLIT DISPLAY
• BAR GRAPH
• LEAD STORAGE
• MAX / MIN
• 3-5/6 DIGIT
6000 COUNT LCD
• AUTO / MANUAL
RANGE
• AUTO HOLD
HOLD
Apo AUTO
RANGE
TRMS
A-HOLD
SYMBOLS USED ON LCD
~
AC Measurement
DC Measurement
-
Negative DC Value
AT
Auto Range Active
O.L.
Overload: Range Exceeded
Apo
Auto Power-Off Active
Low Battery
HOLD
Hold Active
MIN
Minimum Reading
MAX
Maximum Reading
%
Duty Cycle Mode
Frequency Mode
V
Voltage Measurement
A
Current in Amps
Ω
Resistance in Ohms
Diode Test
F
Capacitance in Farads
Continuity Test
Relative / Zero Mode
n
Nano 10
-9
µ
Micro 10
-6
m
Milli 10
-3
k
Kilo 10
3
M
Mega 10
6
°F
Degrees Fahrenheit
°C
Degrees Celsius
A- HOLD
Auto-Hold Active
LOW
Z
Low Impedance
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range
Resolution
Accuracy
600mV ˜ 600V
0.1mV ˜ 0.1V
± (0.2% + 5 digits)
1000V
1V
± (0.5% + 10 digits)
Overload Protection:
1000V
Input Impedance (Nominal)
: >10M
Ω
, <100pF
Input Impedance (LOW Z)
: >3k
Ω
, <200pF, up to 600V RMS
AC Voltage Measurement
Range
Resolution
Accuracy
40Hz ~ 400Hz
Accuracy
400Hz ~ 2kHz
600mV
0.1mV
± (0.75% + 5 digits)
± (2.5% +10 digits)
6V ~ 600V
1mV ~ 0.1V
± (2.0% + 8 digits)
1000V
1V
± (0.75% + 8 digits)
± (2.0% +8 digits)*
Continued on next page.
Overload Protection:
1000V RMS
Input Impedance (Nominal)
:
>10M
Ω
, <100pF
Input Impedance (LOW Z)
: >3k
Ω
, <200pF, up to 600V RMS
Frequency Response:
40 to 2kHz
Response:
True RMS
*Accuracy for 400Hz ~ 1kHz range only.
DC Current Measurement
Range
Resolution
Accuracy
600µA ~ 400mA
0.1µA ~ 0.1mA
± (1.0% + 5 digits)
6A ~ 10A
1mA ~ 10mA
± (1.5% + 5 digits)
Overload Protection: µAmA:
400mA (F440mA/1000V fuse) /
A Input:
10A
(
F11A/1000V fuse)
AC Current Measurement
Range
Resolution
Accuracy
40Hz ~ 2kHz
600µA ~ 400mA
0.1µA ~ 0.1mA
± (1.2% + 5 digits)
6A ~ 10A
1mA ~ 10mA
± (1.8% + 5 digits)
Overload Protection: µAmA:
400mA (F440mA/1000V fuse) /
A Input:
10A
(
F11A/1000V fuse)
Frequency:
40 to 2kHz
Response:
True RMS
Resistance Measurement
Range
Resolution
Accuracy
600
Ω
˜ 6M
Ω
0.1
Ω
˜ 0.001M
Ω
± (1.0% + 5 digits)
50M
Ω
0.01M
Ω
± (1.5% + 10 digits)
Overload Protection:
600V RMS
Capacitance Measurement
Range
Resolution
Accuracy
60nF
0.01nF
± (3.5% + 6 digits)
600nF ~ 6000µF
0.1nF ~ 1µF
Overload Protection:
600V RMS
Frequency Measurement
Range
Resolution
Accuracy
99.99Hz ˜ 399.9kHz
0.01Hz ˜ 0.1kHz
± (0.1% + 3 digits)
Overload Protection:
600V RMS
Sensitivity:
2.5V RMS
Duty Cycle Measurement
Range
Resolution
Accuracy
1.0 ˜ 99.0%
0.1%
± (0.2% per kHz + 0.1% + 8 digits)
Overload Protection:
600V RMS
Sensitivity: < 50V RMS
Frequency Range:
0.5Hz to 10kHz
Diode Test
Overload Protection
Range
Test Current
Open Circuit Voltage
600V RMS
6.0V DC
0.25mA
< 1.8V DC
Continuity Test
Overload Protection
Open Circuit Voltage
Threshold (Appx.)
600V RMS
< 0.44V
<40
Ω
Temperature Measurement
Range
Resolution
Accuracy
-58 ~ 600°F
0.1°F
± (1.0% + 5.4°F)
-50 ~ 600°C
0.1°C
± (1.0% + 3.0°C)
600 ~ 1832°F
1°F
± (1.0% + 5.4°F)
600 ~ 1000°C
1°C
± (1.0% + 3.0°C)
Overload Protection:
600V RMS
K-type
Thermocouple Accuracy:
Not specified
Specifications subject to change.
WARRANTY
See www.kleintools.com, contact Klein Tools at 1-877-775-5346, or visit a
distributor for warranty information.
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using
a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period
of time. Do not expose to high temperatures or humidity. After a period of
storage in extreme conditions exceeding the limits mentioned in the General
Specifications section, allow the instrument to return to normal operating
conditions before using it.
DISPOSAL / RECYCLE
Caution:
This symbol indicates that equipment and its accessories
shall be subject to a separate collection and correct disposal.
CUSTOMER SERVICE
DO NOT apply voltage to thermocouple.
•
Fahrenheit range: -58° to 1832°F (Celsius range: -50° to 1000° C).
8.
Temperature:
Features:
MAX/MIN
REL
HOLD
Testing Fuse:
•
When selecting uA or mA/A functions the meter will indicate
“
LEAd
” if the test leads are not properly connected to the meter.
•
Note: This message remaining on the screen after properly
connecting the test leads is an indication that the fuse is open.
KLEIN TOOLS, INC.
450 Bond Street
Lincolnshire, IL 60069
1-877-775-5346
www.kleintools.com
Rev A 1/13
139750
7.
Frequency (Hz) / Duty Cycle: <400kHz
Features:
HOLD
6.
Low Impedance (LOW Z): < 600V
DO NOT use (LOW Z) mode at voltages greater than 600V.
•
Press SELECT to choose AC or DC Voltage.
•
Use Low Impedance (LOW Z) mode to reduce ghost voltages.
•
See AC/DC Voltage instruction.
Capacitance Features:
HOLD
Safely discharge capacitor before
measurement.
•
Reading may take up to 60 seconds
for large capacitors.
AUTO HOLD
4007036