![KLA Tencor P-16+ Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647084.webp)
Scan Recipes - Creating and Editing a Scan Recipe
KLA-Tencor P-16+ / P-6 User’s Guide
3-36
KLA-Tencor Confidential
0142530-000 AB
3/13/09
Min. Plateau Width
Minimum Plateau Width defines the minimum horizontal distance between rising and
falling edges (or falling and rising edges). This parameter is useful to differentiate
between features of similar shape, but different width. It can also be useful to prevent
false detection of a feature.
Use values between 0.005 and 1000
μ
m.
Apply Gaussian Noise Filter Before Detection
This is used so that feature detection can detect designated features. (See
.) Up edge, up base, down edge, down base, convex, and concave Feature
Detection operate on the first derivative of the trace. As can be seen in the figure,
without any filtering the derivative does not have a clear inflection at the rising and
falling edges of the steps. Since each motion of the stylus up and down due to
roughness or noise is a change in slope, filtering for Feature Detection is critical to
smooth the profile and produce clear derivatives at the rising and falling edges, as can
be seen in the figure after applying a filter.
It does not apply the filter to scan used for parameter calculations. It is only used
for the purpose of Feature Detection.
For use of the
Gaussian Filter
with scan data,
see
NOTE:
It is always recommended to apply a filter to the data. Filtering
can be very useful to obtain robust Feature detection performance.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...