![KLA Tencor P-16+ Скачать руководство пользователя страница 426](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647426.webp)
Preventive Maintenance - Periodic Preventive Maintenance
KLA-Tencor P-16+ / P-6 User’s Guide
15-2
KLA-Tencor Confidential
0142530-000 AB
3/13/09
Monthly PM Procedures
15
Semi-Annual PM Procedures
15
Table 15.2
Monthly PM Procedures
TASK
DESCRIPTION
REASON
RESOURCES
TIME
1.
Perform the
Scan Position Offset - Coarse
Calibration.
To ensure accurate
low-power optics video
positioning with respect
to the stylus.
2.
Check, and if necessary, perform the
Step Height Calibration for 131-
μ
m
range (Contact mode).
To ensure accurate high
range step height
measurements.
3.
Check, and if necessary, perform the
Step Height Calibration for 26-
μ
m
range (Contact mode).
To ensure accurate
Medium Range step
height measurements.
4.
Check, and if necessary, perform the
Step Height Calibration for 6.5-
μ
m
range (Contact mode).
To ensure accurate Low
Range step height
measurements.
5.
Perform the contact version of the
Scan Position Offset - Fine Calibration
To ensure accurate low
power optics video
positioning with respect
to the stylus.
6.
Defragment the hard drive with
Microsoft Defrag Utility.
To ensure efficiency of
computer file system.
Windows
7.
Perform PBackup
To prevent loss of date in
case of an unexpected
failure
See PBackup section in this
manual.
Table 15.3
Semi-Annual PM Procedures
TASK #
DESCRIPTION
REASON
RESOURCES
TIME
1.
Check all cables, connections and
conduit, for binding, pinching or
breakage.
To ensure connectivity of
cables and prevent wear
or damage.
2.
Perform PBackup
To prevent loss of date in
case of an unexpected
failure
See PBackup section in
this manual.
NOTE:
Only KLA-Tencor certified service/maintenance personnel are
authorized to perform these lubrication procedures.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...