830.013.02.K_Installation Guidelines (USA ONLY!) - cBoxX (Version 1.3).docx
10 | P a g e
Metal cable conduits are not allowed for grounding. The piping of the chiller (supply and return)
must also be grounded.
Do not share the ground wire with
other devices.
Always use a ground wire that complies
with technical standards for electrical
equipment and minimizes the length of
the ground wire.
When using more than one Inverter, be careful not to loop the ground.
4.3 Data Cable
:
1” conduit must be provided from the chiller to the MRI control room to allow for pulling the
provided 4 wire cable (50 m long) for connecting the chiller to the remote display panel
The remote display panel controls the complete function of the controller in the main chiller.
5.0 Options and Accessories
5.1 Chiller Interface Panel (CIP)
Weight:
Net (Empty) weight CIP: Approx. 121,25 lbs. (55 kg)
Gross Weight (Operation – Wet): Approx. 132,27 lbs. (60 kg)
Transport weight: Approx. 242,50 lbs. (110 kg).
Dimensions:
Depth: Approx. 14 in. (346 mm).
Width: Approx. 24 in. (610 mm).
Height: Approx. 43 in. (1,100 mm).
Width (Incl. mounting screws): Approx. 25 in. (644 mm).
Height (open valves): Approx. 53 in. (1352 mm).
Clearance:
Maintain at least 20 in. (500 mm) from the top and bottom of the CIP.
Maintain at least 12 in. (300 mm) from the left and right side of the CIP.
Maintain at least 40 in. (1,000mm) from the front of the CIP for servicing.