7. Maintenance
The RaZON
+
itself does not need any maintenance. The following things can be checked on visiting intervals:
- Periodically check the condition of all cables and connectors
- Check if bubble level is still within range
- Check the alignment of the pyrheliometer
- Check the radiometers and clean if required
• Clean pyranometer dome with water or alcohol
• Clean PH1 by unscrewing the sensor part and clean the diffusor with water or alcohol
• Check the PH1 tube for dirt and clean if required
• Remount the sensor part by entering in the slot and rotate until it blocks
The RaZON
+
itself does not need re-calibration. However we suggest having the measuring instruments, used on the RaZON
+
,
re-calibrated every two years.
The PH1 and PR1 have internal desiccant that needs to be replaced every 5 years. When sending the instruments back to Kipp &
Zonen for calibration we will always renew the desiccant.
The label on the radiometers shows the last calibration date. Therefore it also indicates if the desiccant is still active.
The PH1 detector part has a Bayonet connection with the PH1 collimation tube. To open the PH1 rotate the detector part 115 degree
counter clock wise and pull it off.
To close the PH1, push the detector part with a little force in the tube opening and rotate the detector 115 degree clock wise. The
detector will click in position when in place.
Содержание RaZON+
Страница 1: ...RaZON Solar Monitoring System Instruction Manual ...
Страница 2: ...2 ...
Страница 4: ...4 ...
Страница 6: ...6 ...
Страница 12: ...12 ...
Страница 42: ...42 ...
Страница 44: ...44 ...
Страница 50: ...50 ...
Страница 52: ...52 ...
Страница 71: ...71 Example of a logged ASCII data file via Wi Fi Logging Eport Log File ...