Instructions to User 

Thank you for purchasing the Kinetik Wellbeing Finger Pulse Oximeter. 
This  Manual  is  written  and  compiled  in  accordance  with  the

 

council  directive  MDD93/42/EEC  for  medical 

devices and harmonized standards. In case of modifications and software upgrades, the information contained in 
this document is subject to change without notice. 
The  Manual  describes,  in  accordance  with  the  Pulse  Oximeter’s  features  and  requirements,  main  structure, 
functions,  specifications,  correct  methods  for  transportation,  installation,  usage,  operation,  repair,  maintenance 
and storage, etc. as well as the safety procedures to protect both the user and equipment. Refer to the respective 
chapters for details. 
Please read the User Manual carefully before using this product. The User Manual which describes the operating 
procedures  should  be  followed  carefully.  Failure to  follow the  User Manual  may  cause  measuring abnormality, 
equipment damage and human injury. The manufacturer is not responsible for the safety, reliability, performance 
and any monitoring abnormality, human injury, and equipment damage due to users' negligence of the operation 
instructions. The manufacturer’s warranty service does not cover such faults.         
Owing to the forthcoming  renovation, the specific products you received may not be totally in accordance with 
the description of this User Manual. We would sincerely regret for that.                                                                                                                                                             

 

This product is medical device, which can be used repeatedly.   

WARNING: 

 

Uncomfortable  or  painful  feeling  may  appear  if  using  the  device  ceaselessly,  especially  for  the 
microcirculation barrier patients. It is recommended that the sensor should not be applied to the same 
finger for over 2 hours. 

 

For the special patients, there should be a more prudent inspection in the placing process. The device 
cannot be clipped on the edema and tender tissue. 

 

The light (the infrared is invisible) emitted from the device is harmful to the eyes, so users and service 
engineers should not stare at the light. 

 

Users cannot use enamel or other makeup. 

 

Users fingernail cannot be too long. 

 

Please refer to the correlative literature about the clinical restrictions and caution. 

 

This device is not intended for treatment.

 

The User Manual is published by our company. All rights reserved. 

1 Safety 

1.1 Instructions for Safe Operations 

 

Check the main unit and all accessories periodically to make sure that there is no visible damage that may 

affect patient’s safety and monitoring performance about cables and transducers. It is recommended that the 
device should be inspected once a week at least. When there is obvious damage, stop using the monitor. 

 

Necessary  maintenance  must be performed  by qualified  service engineers  only.  Users are  not  permitted  to 
maintain it by themselves. 

 

The oximeter cannot be used together with devices not specified in User’s Manual. Only accessories that are 
appointed or recommend by manufacturer can be used with this device. 

 

This product is calibrated during production. 

1.2 Warnings 

 

Explosive  hazard—DO  NOT  use  the  oximeter  in  an  environment  with  inflammable  gas  such  as  some 
ignitable anesthetic agents. 

 

DO NOT use the oximeter while the user is being measured by MRI and CT. 

 

If allergic to rubber do not use this device. 

 

The disposal  of  scrap  instrument and  accessories  (including battery,  plastic  bags,  foams,  and  paper  boxes) 
should follow the local laws and regulations. 

 

Please check the packing before use to make sure the device and accessories are totally in accordance with 
the packing list, or there is a possibility that the device will work abnormally.   

 

Please don't measure this device with function test paper for the device's related information. 

1.3 Attentions 

 

Keep  the  oximeter  away  from  dust,  vibration,  corrosive  substances,  explosive  materials,  high  temperature 
and moisture. 

 

If the oximeter gets wet, please stop operating it. 

 

When taken from a cold environment to warm or humid environment, please do not use immediately.   

 

DO NOT operate keys on front panel with sharp materials. 

 

High temperature or high-pressure steam disinfection of the oximeter is not permitted. Refer to User Manual 
in the relative chapter for instructions of cleaning and disinfection. 

 

Do not immerge the oximeter in liquid. When it needs cleaning, please wipe its surface with medical alcohol 
by soft material. Do not spray any liquid on the device directly. 

 

When cleaning the device with water, the temperature should be lower than 60℃. 

 

If fingers are too thin or too cold,  this may affect the normal measure of the patients'  SpO

2

 and pulse rate, 

please use a thicker finger such as the thumb and middle finger deeply enough into the probe. 

 

Do not use the device on infant or neonatal patients. 

 

The  product  is  suitable  for  children  above  four  years  old  and  adults  (Weight  should  be  between  15kg  to 
110kg).   

 

The device may not work for all patients. If you are unable to achieve stable readings, discontinue use. 

 

The update period of data is less than 5 seconds, which is changeable according to different individual pulse 
rate. 

 

If  some  abnormal  conditions  appear  on  the  screen  during  test  process,  pull  out  the  finger  and  reinsert  to 
restore normal use. 

 

The device has normal useful life for three years since the first electrified use.   

 

The  hanging  rope  attached  the  product  is  made  from  Non-  allergy  material,  if  you  are  sensitive  to  the 
hanging rope, stop using it. In addition, pay attention to the use of the hanging rope, do not wear around the 
neck to avoid any harm. 

 

The  instrument  does  not  have  a  low-voltage  alarm  function,  it  only  shows  the  low-voltage.  Please  change 
batteries when they run out. 

 

The instrument does not have alarm function. Do not use the device in situations where alarms are required.   

 

Batteries must be removed if the device is going to be stored for more than one month, or else batteries may 
leak. 

 

A flexible circuit connects the two parts of the device. Do not twist or pull on the connection. 

1.4 Indication for Use 

The Fingertip Pulse Oximeter is a non-invasive device intended for the spot-check of oxygen saturation of arterial 
hemoglobin  (SpO

2

)  and  the  pulse  rate  of  adult  and  pediatric  patients  in  home  and  hospital  environments 

(including  clinical  use  in  internist/surgery,  anesthesia,  intensive  care  etc.).  This  device  is  not  intended  for 
continuous monitoring. 

2 Overview 

The pulse oxygen saturation is the percentage of HbO

2

 in the total Hb in the blood, so-called the O

concentration 

in the blood. It is an important bio-parameter for respiration. For the purpose of measuring the SpO

2

 more easily 

and  accurately,  Kinetik  Wellbeing have developed this  Pulse  Oximeter.  This device can  also measure the pulse 
rate simultaneously. 
The  Pulse  Oximeter  includes  features  such  as;  low  power  consumption,  convenient  operation,  small  and  being 
portable. It is only necessary for users to put one of their fingers into a fingertip photoelectric sensor for diagnosis, 
and a display screen will directly show measured value of Hemoglobin Saturation.   

2.1 Classification: 

 

Class II b (MDD93/42/EEC IX Rule 10) 
Class II (U.S.FDA) 

2.2 Features 

 

Operation of the product is simple and convenient. 

 

The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient 
in carrying. 

 

Power  consumption  of  the product is  low, and the two  originally  equipped  AAA  batteries  can  be operated 
continuously for 24 hours. 

 

The product will automatically power off when no signal is in the product within 5 seconds. 

 

Low‐battery indicator as battery icon flash manner. 

2.3 Major Applications and Scope of Application 

The Pulse Oximeter can be used to measure human Hemoglobin Saturation and pulse rate through the finger and 
indicate  the  pulse  intensity  by  the  bar-display.  The  product  is  suitable  for  use  in  family,  Oxygen  bars,  social 
medical organizations and also the measure of saturation oxygen and pulse rate. 
     

  The product is not suitable for use in continuous supervision for patients.

 

        The  problem of  overrating  would  emerge  if  the  patient  is  suffering  from  toxicosis  caused  by  carbon 
monoxide, the device is not recommended to be used under this circumstance.

 

2.4 Environment Requirements 

Storage Environment 
  a) Temperature: -40℃~+60℃ 
  b) Relative humidity: 

95% 

  c) Atmospheric pressure :500hPa~1060hPa 

 
Operating Environment 
  a) Temperature: :10℃~40℃ 
  b) Relative Humidity :≤75% 
  c) Atmospheric pressure:700hPa~1060hPa   

3 Principle and Caution 

3.1 Principle of Measurement 

Principle  of  the  Oximeter  is  as  follows:  An  experience  formula  of  data  process  is  established  taking  use  of 
Lambert-Beer  Law  according  to  Spectrum  Absorption  Characteristics  of  Reductive  Hemoglobin  (Hb)  and 
Oxyhemoglobin  (HbO

2

)  in  glow  &  near-infrared  zones.  Operation  principle  of  the  instrument  is:  Photoelectric 

Oxyhemoglobin  Inspection  Technology  is  adopted  in  accordance  with  Capacity  Pulse  Scanning  &  Recording 
Technology,  so  that  two  beams  of  different  wavelength  of  lights  can  be  focused  onto  human  nail  tip  through 
perspective  clamp  finger-type  sensor.  Then  measured  signal  can  be  obtained  by  a  photosensitive  element, 
information  acquired  through  which  will  be  shown  on  screen  through  treatment  in  electronic  circuits  and 
microprocessor. 

 
 
 
 
 
 

 

 

Figure 1. Operating Principle 

3.2 Caution 

1.

 

The finger should be placed properly (see the attached illustration of this manual, Figure 5), or else it may 
cause inaccurate measurement. 

2.

 

The SpO

2

 sensor and photoelectric receiving tube should be arranged in a way with the subject’s arteriole in 

correct position. 

3.

 

The  SpO

2

  sensor  should not be used at  a  location or limb that is tied  with  arterial  canal or blood  pressure 

cuff or receiving intravenous injection. 

4.

 

Make sure the optical path is free from any optical obstacles like rubberized fabric. 

5.

 

Excessive  ambient  light  may  affect  the  measuring  result.  It  includes  fluorescent  lamp,  dual  ruby  light, 
infrared heater, direct sunlight etc. 

6.

 

Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy. 

7.

 

Users cannot use enamel or other makeup. 

3.3 Clinical Restrictions 

1.

 

As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject is required. 
For  a  subject  with  weak  pulse  due  to  shock,  low  ambient/body  temperature,  major  bleeding,  or  use  of 
vascular contracting drug, the SpO

2

 waveform (PLETH) will decrease. In this case, the measurement will be 

more sensitive to interference. 

2.

 

For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green and acid 
indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, 
and some with icterus problem, the SpO

2

 determination by this monitor may be inaccurate. 

3.

 

The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also cause serious error of SpO

2

 

measurement.   

4.

 

As  the  SpO

2

  value  serves  as  a  reference  value  for  judgement  of  anemic  anoxia  and  toxic  anoxia,  some 

patients with serious anemia may also report good SpO

2

 measurement. 

4 Technical Specifications 

1)

 

Display Format:

 Digital tube Display. 

 

 

SpO

2

 Measuring Range: 

0% - 100%. 

 

Pulse Rate Measuring Range:

 30 bpm - 250 bpm. 

 

Pulse Intensity Display:

 columniation display 

2)

 

Power Requirements:

 2 ×1.5V AAA alkaline battery, adaptable range: 2.6V~3.6V. 

3)

 

Power Consumption:

 Smaller than 25 mA. 

4)

 

Resolution:

 1% for SpO

2

 and 1 bpm for Pulse Rate. 

5)

 

Measurement  Accuracy:

  ±2%  in  stage  of  70%-100%  SpO

2

,  and  meaningless  when  stage  being  smaller 

than 70%. ±2 bpm or±2% (select larger) for Pulse Rate.   
6)

 

Measurement  Performance  in  Weak  Filling  Condition:

  SpO

2

  and  pulse  rate  can  be  shown  correctly 

when pulse-filling ratio is 0.4%. SpO

2

 error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger). 

7)

 

Resistance  to  surrounding  light:

  The  deviation  between  the  value  measured  in  the  condition  of 

man-made light or indoor natural light and that of darkroom is less than ±1%. 
8)

 

It is equipped with a function switch. The Oximeter can be powered off in case no finger is the   

9)     

Optical Sensor 

          Red light (wavelength is 660nm, 6.65mW)   
          Infrared (wavelength is 880nm, 6.75mW)     

5 Accessories 

 

One hanging rope. 

 

Two batteries. 

 

One User Manual. 

6 Installation 

6.1 View of the Front Panel 

                 

 

      Figure 2. Front View                                                        Figure 3. Batteries Installation

 

6.2 Battery 

Step 1. Refer to Figure 3. and insert the two AAA size batteries properly in the right direction. 
Step 2. Replace the cover.   

Please take care when you insert the batteries as improper insertion may damage the device. 

 

6.3 Mounting the Hanging Rope   

Step 1. Put the end of the rope through the hole. 
Step 2. Put another end of the rope through the first one and then tighten it. 

           

                                         

 

Figure 4. Mounting the hanging rope                                          Figure 5. Put finger in position 

7 Operating Guide 

7.1 Insert the two batteries properly to the direction, and then replace the cover. 
7.2 Open the clip as shown in Figure 5.   
7.3 Let the patient’s finger put into the rubber cushions of the clip (make sure the finger is in the right position), 
and then clip the finger. 
7.4 Press the switch button once on front panel. 
7.5 Do not shake the finger and keep the patient at ease during the process. Measurement is not recommended if 
moving.   
7.6 Get the information directly from screen display. 
7.7 In boot-strap state, press button, and the device is reset. 

  Fingernails and the luminescent tube should be on the same side.

 

8 Repairing and Maintenance 

 

Please change the batteries when the low voltage symbol is displayed on the screen. 

 

Please clean the surface of the device before using. Wipe the device with medical alcohol first, and then let 

it dry in air or clean it by dry clean fabric.   

 

Use medical alcohol to disinfect the product after use, to prevent from cross infection for next use. 

 

Please take out the batteries if the oximeter is not in use for a long time. 

 

The best storage environment of the device is - 40ºC to 60ºC ambient temperature and not higher than 95% 

relative humidity.   

 

Users  are  advised  to  calibrate  the  device  termly  (or  according  to  the  calibrating  program  of  hospital).  It 

also can be performed at the state-appointed agent or just contact us for calibration. 

  High-pressure sterilization cannot be used on the device.   