Password Protection
The system must be enabled to allow access to other menus.
The enable command is “enable12345”.
Operate Page Commands
Epi2.0 Operate Menu:OPERATE> ?
Commands:
4G: Sets Sensors to 4g FS Range
2G: Sets Sensors to 2g FS Range
1G: Sets Sensors to 1g FS Range
0.5G: Sets Sensors to 0.5g FS Range
0.25G: Sets Sensors to 0.25g FS Range
AUTOZEROON: Turns ON Autozero Mode
AUTOZEROOFF: Turns OFF Autozero Mode
CALON: Enables Sensor CAL
CALOFF: Disables Sensor CAL
STATUS: Prints Sensor Operating Mode Values
SAFE: Disables Sensor System Control
RETURN: Return to Previous Menu
?: Help for info on specific command
Seconds=4302
Ticks=369
Epi2.0 Operate Menu:OPERATE>
The “status” command prompts the system to print out a set of system-related data:
Epi2.0 Operate Menu:OPERATE> status
Sensor FS Range is 4g
AUTOZERO is OFF
CAL is OFF
Seconds=4364
Ticks=36
Epi2.0 Operate Menu:OPERATE>
Page 17